TDS Meters
Bante530 Portable Conductivity/TDS Meter
Professional portable TDS meter with conductivity measurement mode, 1 to 5 points calibration, setup menu contains 12 options, accuracy: 1% F.S.
Features:
• Professional portable conductivity/TDS meter is equipped with a backlit LCD display.
• 1 to 5 points calibration with automatic recognition for conductivity standards.
• Selectable cell constant (K=0.1/1/10), reference temperature (20/25°C), TDS factor, linear and
pure water compensation modes.
• Automatic temperature compensation corrects conductivity measurements to the reference
temperature.
• Auto-read function senses and locks the measurement endpoint.
• Calibration due alarm reminds the user to calibrate the meter regularly.
• Calibration log shows the calibration points and factor.
• Setup menu allows setting the number of calibration points, stability criteria, temperature unit,
date and time, auto-power off, etc.
• Reset function automatically resumes all settings back to the factory defaults.
• Expanded memory stores or recalls up to 500 data sets.
• USB communication interface for data transfer or connecting a power adapter to the meter.
Model | Bante530 | |
Conductivity | ||
Range | 0 to 20.00, 200.0, 2000 µS/cm, 20.00, 200.0 mS/cm | |
Resolution | 0.001, 0.01, 0.1, 1, automatic | |
Accuracy | ±0.5% F.S. | |
Calibration | 1 to 5 points | |
Calibration Solutions | 10 µS / 84 µS / 1413 µS / 12.88 mS / 111.8 mS | |
TDS | ||
Range | 0 to 10.00, 100.0, 1000 ppm, 10.00, 200.0 ppt | |
Resolution | 0.01, 0.1, 1, automatic | |
Accuracy | ±1% F.S. | |
TDS Conversion Factor |
| 0.1 to 1.0 (default 0.5) |
Temperature | ||
Range | 0 to 105°C / 32 to 221°F | |
Resolution | 0.1°C/°F | |
Accuracy | ±0.5°C / ±0.9°F | |
Calibration | 1 point | |
Calibration Range | Reading ±10°C/°F | |
Other Specifications | ||
Temperature Compensation | 0 to 100°C / 32 to 212°F, manual or automatic | |
Temperature Coefficient |
| 0.0 to 10.0%/°C |
Pure Water Compensation |
| Yes |
Reference Temperature |
| 20 / 25°C |
Cell Constant |
| K=0.1 / 1 / 10 / Custom |
Hold Function |
| Manual / Auto-read measurement endpoint |
Stability Criteria |
| Low / High |
Calibration Due Alarm |
| 1 to 31 days or off |
Auto-Power Off |
| 10 / 20 / 30 minutes after last key pressed |
Memory |
| Stores up to 500 data sets |
Output |
| USB communication interface |
Connector |
| 6-pin mini-DIN, 3.5 mm jack socket |
Display |
| Custom LCD (80 × 60 mm) |
Power Supply |
| 3 × 1.5V AA batteries or 5V DC power adapter |
Battery Life | Approximately 150 hours without backlight | |
Environmental Conditions | 0 to 50°C / 32 to 122°F, relative humidity < 80% | |
Dimensions | 170 × 85 × 30 mm | |
Weight | 300 g |
Ordering Code | Meter Includes |
Bante530-S | CON-1 conductivity electrode |
TP-10K temperature probe | |
Conductivity standard solutions 84 µS/cm, 1413 µS/cm, 12.88 mS/cm | |
Electrode clip | |
Carrying case | |
Bante530-DL | CON-0.1 conductivity electrode |
CON-1 conductivity electrode | |
TP-10K temperature probe | |
Conductivity standard solutions 84 µS/cm, 1413 µS/cm, 12.88 mS/cm | |
Electrode clip | |
Carrying case | |
Bante530-DH | CON-1 conductivity electrode |
CON-10 conductivity electrode | |
TP-10K temperature probe | |
Conductivity standard solutions 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm | |
Electrode clip | |
Carrying case |
Items | Description |
CON-0.1 Conductivity Electrode | For measuring the pure water |
CON-1 Conductivity Electrode | For measuring the general water samples |
CON-10 Conductivity Electrode | For measuring the high conductivity samples |
TP-10K Temperature Probe | For temperature compensation and measurement |
Conductivity Standard Solutions | For conductivity calibration |
USB Cable | For transferring data or connecting power adapter |
5V DC Power Adapter | For connecting external power source |
Ordering Code | Dimensions and Weights |
Bante530-S | 35 × 10 × 30 cm, 1.5 kg |
Bante530-DL | 35 × 10 × 30 cm, 1.5 kg |
Bante530-DH | 35 × 10 × 30 cm, 1.5 kg |