Bante210 Benchtop pH Meter

Classical benchtop pH meter, 1 to 3 points calibration, automatic temperature compensation, setup menu contains 5 options, accuracy: 0.01 pH.

Features:
• Classical benchtop pH meter is equipped with a backlit LCD display.
• 1 to 3 points calibration with automatic recognition for USA and NIST buffers.
• Automatic electrode diagnosis helps the user determine whether to replace the pH electrode.
• Automatic temperature compensation ensures accurate readings over the entire range.
• Auto-read function senses and locks the measurement endpoint.
• Setup menu allows setting the pH buffer set, number of calibration points, temperature unit,
  auto-power off, etc.
• Reset function automatically resumes all settings back to the factory defaults.

Benchtop pH Meter

Applications:
Education: Chemical experiments
Laboratory: Routine measurements and quality controls

   Model


Bante210

   pH



   Range


-1.00 to 15.00 pH

   Resolution


0.01 pH

   Accuracy


±0.01 pH

   Calibration


1 to 3 points

   pH Buffer Options


USA (pH 4.01/7.00/10.01) / NIST (pH 4.01/6.86/9.18)

   mV



   Range


-1999 to 1999 mV

   Resolution


1 mV

   Accuracy


±1 mV

   Temperature



   Range


0 to 105°C / 32 to 221°F

   Resolution


0.1°C/°F

   Accuracy


±1°C / ±1.8°F

   Calibration


1 point

   Calibration Range


Reading ±10°C/°F

   Other Specifications



   Temperature Compensation


0 to 100°C / 32 to 212°F, manual or automatic

   Hold Function


Manual / Auto-read measurement endpoint

   Connector

BNC, 3.5 mm jack socket

   Display

Custom LCD (135 × 75 mm)

   Power Supply

9V DC power adapter

   Environmental Conditions

0 to 50°C / 32 to 122°F, relative humidity < 80%

   Dimensions


210 × 205 × 75 mm

   Weight


1.5 kg

   Ordering Code

Meter Includes

   Bante210-CN

E201-BNC pH electrode

TP-10K temperature probe

pH 4.01, 7.00, 10.01 buffer reagents

9V DC power adapter

   Bante210-UK

   

P11 glass pH electrode

TP-10K temperature probe

pH 4.01, 7.00, 10.01 buffer reagents

9V DC power adapter

Items Description
E201-BNC pH Electrode For measuring the liquids
E202-BNC pH Electrode For measuring the semisolid and gel samples
P11 pH Electrode For measuring the weak corrosive samples
P11-LiCl pH Electrode For measuring the non-aqueous samples
P11-HA pH Electrode For measuring the high alkalinity samples
P11-NA pH Electrode For measuring the biofuels
P13 pH Electrode For measuring the micro-volume samples
P15 pH Electrode For measuring the low conductivity samples
P16 pH Electrode For measuring the Tris buffers
P19 pH Electrode For measuring the semisolid samples
P22 pH Electrode For measuring the high temperature samples
TP-10K Temperature Probe For temperature compensation and measurement
pH Buffer Solutions For pH calibration
pH Electrode Storage Solution For electrode maintenance and storage
pH Electrode Cleaning Solution For general cleaning
pH Electrode Cleaning Solution For removing the protein contamination

   Ordering Code

Dimensions and Weights

   Bante210-CN

30 × 25 × 20 cm, 1.7 kg

   Bante210-UK

30 × 25 × 20 cm, 1.7 kg