Multiparameter Meters

Bante901 Benchtop pH/Conductivity Meter

High-precision benchtop pH conductivity meter provides accurate measurements of pH, mV, conductivity and TDS, suitable for laboratory applications.

Features:
pH
• 1 to 5 points calibration with automatic recognition for USA, NIST and DIN buffers.
• Automatic electrode diagnosis shows the pH slope and zero offset.

pH Conductivity Meter

Conductivity/TDS
• 1 to 5 points calibration with automatic recognition for conductivity standards.
• Automatic electrode diagnosis shows the calibration points and factor.
• Selectable cell constant (K=0.1/1/10), reference temperature (20/25°C), TDS factor, linear and
  pure water compensation modes.

Other Features
• Automatic temperature compensation ensures accurate readings over the entire range.
• Auto-read function senses and locks the measurement endpoint.
• Calibration due alarm reminds user to calibrate the meter regularly.

pH Conductivity Meter

• Setup menu allows setting the number of calibration points, stability criteria, temperature units,
  date and time, auto-power off, etc.
• Reset function automatically restores all settings to the factory defaults.
• Expanded memory stores or recalls up to 500 data sets.
• USB communication interface enables easy data transfer to a PC.
• DAS software is capable of receiving data, recording measurements at specified intervals,
  exporting data to Excel.

   Model


Bante901

   pH



   Range


-2.000 to 20.000 pH

   Resolution


0.001 / 0.01 / 0.1, selectable

   Accuracy


±0.002 pH

   Calibration


1 to 5 points

   pH Buffer Options


USA / NIST / DIN / 2 custom buffers

   mV



   Range


-1999.9 to 1999.9 mV

   Resolution


0.1 / 1, selectable

   Accuracy


±0.2 mV

   Conductivity



   Range


0.01 µS/cm to 200.0 mS/cm

   Resolution


0.001, 0.01, 0.1, 1, automatic

   Accuracy


±0.5% F.S.

   Calibration


1 to 5 points

   Calibration Solutions


10 µS / 84 µS / 1413 µS / 12.88 mS / 111.8 mS

   Temperature Coefficient


0.0 to 10.0%/°C

   Pure Water Compensation


Yes

   Reference Temperature


20 / 25°C

   Cell Constant  


K=0.1 / 1 / 10 / Custom

   TDS



   Range


0.01 ppm to 200.0 ppt

   Resolution


0.01, 0.1, 1, automatic

   Accuracy


±1% F.S.

   TDS Conversion Factor


0.1 to 1.0 (default 0.5)

   Temperature



   Range


0 to 105°C / 32 to 221°F

   Resolution


0.1

   Accuracy


±0.5°C / ±0.9°F

   Calibration


1 point

   Calibration Range


Reading ±10°C/°F

   Other Specifications



   Temperature Compensation


0 to 100°C / 32 to 212°F, manual or automatic

   Hold Function


Manual / Auto-read measurement endpoint

   Stability Criteria


Low / High

   Calibration Due Alarm


1 to 31 days or off

   Memory


Stores up to 500 data sets

   Output


USB communication interface

   Connector


BNC, 6-pin mini-DIN, 3.5 mm jack socket

   Display


Custom LCD (125 × 100 mm)

   Power Supply


5V DC power adapter

   Environmental Conditions


0 to 50°C / 32 to 122°F, relative humidity < 80%

   Dimensions


210 × 188 × 60 mm

   Weight


1.5 kg

   Ordering Code

Meter Includes

   Bante901-CN

E201-BNC pH electrode

CON-1 conductivity electrode

TP-10K temperature probe

pH 4.01, 7.00, 10.01 buffer reagents

Conductivity standard solutions 84 µS/cm, 1413 µS/cm, 12.88 mS/cm

USB cable

5V DC power adapter

   Bante901-UK

   

P11 glass pH electrode

K10 conductivity electrode

TP-10K temperature probe

pH 4.01, 7.00, 10.01 buffer reagents

Conductivity standard solutions 84 µS/cm, 1413 µS/cm, 12.88 mS/cm

USB cable

5V DC power adapter

Items Description
E201-BNC pH Electrode For measuring general samples
E202-BNC pH Electrode For measuring semisolid and gel samples
P11 pH Electrode For measuring mildly corrosive samples
P11-LiCl pH Electrode For measuring non-aqueous samples
P11-HA pH Electrode For measuring high alkalinity samples
P13 pH Electrode For measuring micro-volume samples
P15 pH Electrode For measuring low conductivity samples
P16 pH Electrode For measuring Tris-buffered samples
P19 pH Electrode For piercing and measuring soft solids
CON-0.1 Conductivity Electrode For measuring low conductivity samples
CON-1 Conductivity Electrode For measuring general water samples
CON-10 Conductivity Electrode For measuring high conductivity samples
pH Buffer Solutions For calibrating pH electrode
Conductivity Standard Solutions For calibrating conductivity electrode

   Ordering Code

Dimensions and Weights

   Bante901-CN

30 × 25 × 20 cm, 1.7 kg

   Bante901-UK